Model
Z-10
Function
Illuminance(lux)、Illuminance ratio(%)、
Integrated illuminance(lux·h)、
Average illuminance(lux)
Detector
Silicon optical device
V(λ) matching error
f1'≤3.5%( standard class)
or f1'≤5%(class 1)
Illuminance
Range
Lux:0.01lux~300klux;
lux·h:0.01lx·h~999.99Mlx·h
Accuracy
±4%(according to JJG245-2005)
or ±2%(according to JJG245-2005)
Dispaly
5-digit backlit LCD display
Communication
USB
User Calibration
CCF color correction coefficient function,
zero setting, calibration