Excellent long-term stability: The resistance elements are made of selected manganin wire, stabilized by both artificial and natural aging after being wound.
Perfect guard system
Stable measurement: The dial switches offer low contact resistance and little change due to aging, giving stable operation. Inner/outer contact type plugs are used for changing the multiplication factor, assuring little change in the resistance due to reinsertion of plugs.
Laboratory-grade accuracy ±0.03 to ±0.05%
Quick readout without error
Built-in standard resistors
Bar resistor measurement
Thông số kỹ thuật
Dimensions |
227(W)×183(H)×93(D)mm, excluding protrusions
8.94 × 7.20 × 3.66 in. |
Weight |
Approx. 2.6kg
5.7 lbs |
Measuring Range |
0.1mΩ to 110Ω |
Measuring Arm |
1.00 to 11.00Ω at x1 |
Multiplier |
0.005mΩ at x 0.0001
0.05mΩ at x 0.001
0.5Ω at x 10
0.5mΩ at x 0.01
50mΩ at x 1
5mΩ at x 0.1
x 0.0001, x 0.001, x 0.01, x 0.1, x 1, x 10 (plug-in system) |
Accuracy |
±(0.05Ω x multiplier +0.01mΩ) at temperature of 5 to 35°C (41 to 95°F) and humidity range of less then 85% |
Galvanometer (Built-In) |
Galvanometer (built-in) electronic DC galvanometer, Voltage Sensitivity approx. 20μV/div.
Sensitivity Changeover
G0 - Approx. 11kΩ (input Resistance)
G1 - Approx. 1/11 of G0 Sensitivity
G2 - Approx. 1/11o of G0 Sensitivity
Power Source:1 9V Battery (JIS6F22)
Battery life ;Approx. 300 Hours |
Operating Temperature Range |
5 to 35°C (41 to 95°F) |
Humidity Range |
85% Max, Relative Humidity |
Accessories Supplied at No Extra Cost |
Carrying Case |
Optional Accessories |
2753 Measuring Cords
2754 Clamp Device
2771 Standard Resistor |
Note |
Standard Resistor (Yokogawa 2771) Is required for measurements on 0.1 to 1.1mΩ range at 0.0001 multiplier |