3 windows display
|
V、A、W、VA、var、PF、Hz、kWh、time、deg、ORD、%、CF
|
Input Voltage
|
600V/300V/150V
|
Input Current
|
50A/10A/2A
|
Range switch
|
AUTO
|
Frequency range
|
AC:15Hz~400Hz
|
Measurement Accuracy(sine wave)
|
Voltage、Current
|
Power
|
15Hz~45Hz
|
±(0.6%Reading +0.4%Range+1digit)
|
±(1.5%Reading + 0.5%Range+1digit)
|
45Hz~65Hz
|
±(0.4%Reading +0.1%Range+1 digit)
|
±(0.8%Reading + 0.2%Range+1 digit)
|
65Hz~400Hz
|
±(0.8%Reading +0.2%Range+1 digit)
|
±(1.5%Reading + 0.5%Range+1 digit)
|
Measurement Accuracy
(PWM waveform)
|
30Hz~45Hz
|
±(3%Reading + 2%Range+1 digit)
|
±(2%Reading + 3%Range+1 digit)
|
45Hz~65Hz
|
±(1%Reading + 1%Range+1 digit)
|
±(2%Reading + 1%Range+1 digit)
|
65Hz~400Hz
|
±(1%Reading + 2%Range+1 digit)
|
±(2%Reading + 2%Range+1 digit)
|
Harmonic analysis
|
Measure the THD and 2nd -50th sub-harmonic of voltage and current
|
Integration mode
|
Standard integration mode, Manual integration mode and Continuous integral mode.
|
Integrating time
|
00.00.00~99.59.59
|
Electrical integrating time
|
-99999kWh~999999kWh
|
Scaling function
|
External transformer, voltage / current / power ratio coefficient can be set.
|
Communication mode
|
RS-232 interface
|