Model |
HC-200F |
Combined sensor |
Fluoride ion electrode 1009 |
Measurement range |
Concentration:
0 mg/L to 10000 mg/L (Display range: 0 mg/L to 11000 mg/L)
The selectable range: 10000 ,5000 ,2000 ,1000 ,500 ,200 ,100 ,50 ,20.0 ,10.0 mg/L
Temperature: 0°C to 100°C (Display range: -10°C to 110°C ) |
Transmission output |
Number of output: 2
(The negative terminals for transmission outputs are internally connected at the same electric potential)
Output type: 4 mA to 20 mA DC: input/output isolated type
Load resistance: Maximum: 900 Ω |
Contact output |
Number of output: 3
Output type: No-voltage contact output
Contact type: Relay contact; SPDT (1c)
Output capacity: 250 VAC 3 A, 30 VDC 3 A (resistance load) |
Cleaning output |
Number of output: 1
Output type: AC power control output ( Applied power supply voltage)
Contact type: Relay contact; SPST (1a)
Output contact capacity: 250 VAC 3 A, 30 VDC 3 A (resistance load)
External instrument: Solenoid valve for cleaning control |
Contact input |
Number of inputs: 1
Contact type: No-voltage "a" contact for open collector |
Communication function |
Communication type: RS-485
Signal type:
2 wire system, isolated from the input circuit
Not isolated from transmission circuit |