DCX-16
The sensor, electronics and battery are accommodated in one housing. To extract data, the data collector must be removed from the measuring point, and the watertight screw cap that allows access to the read-out plug /interface must be unscrewed. The DCX-16 operates with an absolute pressure sensor. In shallow water, when the influence of air pressure fluctuations has to be taken into account, a second logger (barometric logger, obtainable separately) positioned on the surface of the water must be used to record the air pressure progression. The computer software then calculates the differential pressure or the water level by subtracting the two measured data.
DCX-16 SG/VG
Instead of a watertight screwed closure, these versions have a cable output. The interface plug is fixed on the sounding tube above the surface of the water with the help of a screw-on fixing disk. This means that there is no need to remove the DCX-16 SG/VG from the sounding tube in order to read the data.
In the VG version (reference pressure measurement), the reference pressure-compensating capillary is routed in the same cable into the read-out plug housing; this also contains the reference aperture (protected by a Gore-Tex® diaphragm) which produces the pressure equalization.
|