AC Current Output
|
6-phase (RMS) |
6*0~30A |
| 3-phase (RMS) |
3*0~60A |
| 6-phase parallel current output (RMS) |
0~180A |
| Allowable long time working value of phase current (RMS) |
10A |
| Maximum output power of phase current |
400VA |
| Allowable working hours @ max. output of 6-phase parallel current |
10s |
| Accuracy |
≤0.5A ±2mA
>0.5A 0.2% |
| Frequency range (Fundamental wave) |
0~1000Hz |
| Harmonic times |
0~20 times |
DC Current Output
|
Current output |
0~±10A/phase; 0~±60A/6-pahse parallel |
| Accuracy |
0.5% |
| Maximum output load voltage |
20V |
AC Voltage Output
|
Phase voltage output (RMS) |
0~120V Accuracy: 0.2% |
| Line voltage output (RMS) |
0~240V |
| Phase voltage/Line voltage output power |
80VA/100VA |
| Frequency range (Fundamental wave) |
0~1000Hz |
| Harmonic times |
1~20 times |
DC Voltage Output
|
Phase voltage output amplitude |
0~±160V Accuracy: 0.5% |
| Line voltage output amplitude |
0~±320V |
| Phase voltage/Line voltage output power |
70VA/140VA |
Binary
|
Binary input |
10 channels |
| Idle contact |
1~20mA, 24V |
| Potential contact access |
"0": 0 ~+6V; "1": +11 V ~+250 V |
| Binary output |
8 pairs DC: 220 V/0.2 A; AC: 220 V/0.5 A |
| Time Measuring Range |
0.1ms~9999s Measuring accuracy: <0.1mS |
| Dimension & Weight |
480*360*200mm³ 19kg |