MEASUREMENT DATA |
|
SCALES |
Refractive index (nD)
Concentration of sucrose, glucose, fructose and invert sugar [%Brix],
User-defined |
MEASUREMENT RANGE |
nD 1.3200–1.5800
0–95 %Brix |
nD 1.3200–1.7000
0–95 %Brix |
nD 1.32000–1.58000
0–95 %Brix |
nD 1.32000–1.70000
0–95 %Brix |
MEASUREMENT ACCURACY |
nD ±0.0001
±0.1 %Brix |
nD ±0.00002
±0.02 %Brix |
RESOLUTION |
nD 0.0001
0.1 %Brix |
nD 0.00001
0.01 %Brix |
MEASUREMENT PERIOD |
approx. 4 s |
MEASUREMENT PRISM |
Sapphire |
LIGHT SOURCE |
LED |
WAVELENGTH |
589 nm |
TEMPERATURE DATA |
|
TEMPERATURE CONTROL |
With integrated Peltier temperature control |
TEMP. CONTROL RANGE |
10–80 °C |
ACCURACY OF TEMP. CONTROL |
±0.1 °C |
TEMPERATURE COMPENSATION |
Can be activated
(ICUMSA or freely definable) |
TEMPERATURE MEASUREMENT |
With integrated temperature sensor Pt100 |
TEMPERATURE MEASUREMENT RANGE |
10–80 °C |
TEMP. MEASUREMENT ACCURACY |
±0.1 °C |
TEMP. MEASUREMENT RESOLUTION |
0.1 °C |