Measurement Section |
Range |
0 to 999.9mΩ |
Resolution |
0.1μΩ - below 1.0mΩ
1μΩ - below 10mΩ
10μΩ - below 100mΩ
100μΩ - below 1000mΩ |
Current Shunt |
600A, 60mV |
Sense Ranges |
0 to 2mV, 0 to 20mV, 0 to 200mV, 0-5V |
Inaccuracy |
Typical |
Maximum |
100A, ta 25°C, R < 1mΩ |
±0.2μΩ |
±1μΩ |
50-600A, ta 10-40°C, R < 1mΩ |
±0.3μΩ |
±2μΩ |
50-600A, ta 0 to 50°C, R < 1mΩ |
±0.7μΩ |
±3μΩ |
50 to 600A, ta -20 to +50°C, R < 1mΩ |
±1.1μΩ |
±4μΩ |
600A, ta 10 to 40°C, 1mΩ < R < 8.4mΩ |
±6μΩ |
±50μΩ |
50A, ta 10 to 40°C, 10mΩ < R < 100mΩ |
±80μΩ |
±500μΩ |
5A, ta 10 to 40°C, 100mΩ < R < 1000mΩ |
±1mΩ |
±10mΩ |
Outputs |
DC+/COM |
Range |
5 to 600A DC (steps of 1A) |
Max Output Voltage |
5.25V at 600A |
Max Ripple |
80mVpp, 28.3mVrms at 0 to 50°C (+32 to +122°F) |
Max Load Capacity |
300A (continuous) |
Output 100μV/A |
Shunt Output |
From internal shunt 60mV at 600A |
Inaccuracy |
±1% |
Inputs |
Sense |
Max 20V between terminals and to protective earth (ground) |
Input |
Max 20V between terminals and to protective earth (ground) |
DC Current Clamp Input Sensitivity |
Adjustable from 0.1 to 20mV/A |
DC Current Clamp Input Impedance |
>1MΩ |