Test currents
Up to 100 A (True DC).
The test current can be set between 10% and 100% of the working stream with a resolution of 1 A.
Resistence rangers
0.1 µΩ to 300 mΩ.
Measurement Principle
Four-terminal, Kelvin-type.
Thermal protection
Protects all sensitive components, avoiding any damage due to overheating.
Test with programmable time
Allows programming the test period of 5 seconds to 120 seconds.
Basic accuracy
± 1% of reading ± 1 digits.
Advanced features
Digital direct reading of resistance measured in an alphanumeric LCD display 4 lines / 20 characters. Measurements are obtained quickly and with great accuracy.
Built-in printer
allows print the measured values.
Built-in memory
Capacity to store 50 records with 80 readings each.
Serial data output
USB.
MegaLogg2 Software
Friendly, easy to use software. Tests are represented in graphic and tabular views. With automatic report generator, including the operator's commentaries.
Enviromental
IP54 with closed lid.
Safety class
Meets the requirements of IEC 61010-1.
Power supply
Mains supply: 90-240 V~ 50/60 Hz auto.
Operating temperature range
-5 °C up to +50 °C.
Storage temperature range
-25 °C up to +70 °C.
Humidity range
95% UR (non condensing).
Weigth
Approx. 11 kg.
Dimension
502 x 394 x 190 mm.