| Model: PRS-801B Specifications |
| Resistance |
0.01 Ohms to 200Tera-Ohms |
| Open Circuit Current (I) |
<4ma @ 100Volts |
| Ohms Display indicators |
Ω, KΩ, MΩ, GΩ and TΩ. Includes 19-segment analog scale (1-10 with 0.5 indication) with x1, x10, & x100 multipliers. |
| Number of Data Points in Memory |
0 - 999 |
| Test Voltage |
| Automatic Mode |
0.01Volts to 10Volts Variable: 1.0E-1Ohms to 1.0E+4Ohms |
| Constant Voltage |
10Volts (+/- <0.2Volts): 1.0E+4Ohms to <1.0E+6 Ohms
100Volts (+/- <2.0Volts): 1.0E+6Ohms to 2.0E+14Ohms |
| Manual Mode |
0.01Volts to 10Volts Variable: 1.0E-1Ohms to <1.0E+5Ohms
10Volts (+/- <0.2 volts): 1.0E+2Ohms to 1.0E+9Ohms
100Volts (+/- <2.0 volts): 2.0E+5 to 2.0E+14 Ohms |
| Accuracy+ |
| Overall |
+/- <5% at ambient conditions (at 23°C and 30% Rh) |
| Nominal Range Tolerances |
1.00E-2Ω to 1.00E+1Ω: +/- 10% corrected for test lead resistance
1.10E+1Ω to 9.99E+10Ω: +/- 2.0%
1.00E+11Ω to 9.99E+11Ω: +/- 5.0%
1.00E+12Ω to 2.0E+14Ω: <20% with grounded, shielded lead |
| Electrification |
| Resistance Range |
Electrification Period |
| 0.01 to <1.00E+06 Ohms |
3.0 seconds |
| 1.00E+06 to <1.00E+12 |
8.0 seconds |
| 1.00E+12 and greater |
180 seconds |