Model
|
YL4560
|
Illumination system
|
45/0 method(45 ring-shaped illumination, vertical viewing);Conforms to CIE No.15,GB/T 3978,GB 2893,GB/T 18833,ISO7724-1,ASTM E1164,DIN5033 Teil7,GB 2893,GB/T 18833
|
Features
|
45/0 non- contact benchtop spectrophotometer can achieve precise non- contact measurement between the measuring probe and the sample like liquids, sauces, powders, etc. It can be applied for accurate color analysis and transmission In laboratories and R & D industries; It can also be used for accurate color measurement and quality control of automated production lines; It is widely used in cosmetics, fruits and vegetables, food hygiene, plastic electronics, paint and ink, printing, ceramics and other industries, and also for fluorescence sample measurement.
|
Light Source
|
Full spectrum LED light,UV light
|
Spectrophotometric Mode
|
Concave Grating
|
Sensor
|
256 Image Element Double Array CMOS Image Sensor
|
Wavelength Range
|
400~700nm/10nm output
|
Reflectance Range
|
0~200%
|
Measuring Aperture
|
Φ20mm(can be customized <20mm)
|
Non-contact Distance
|
7.5mm
|
Sample Height
|
0~150mm
|
Distance Adjustment Method
|
Manual adjustment and automatic adjustment (test height can be stored)
|
Measuring mode
|
Sample mode/QC Mode/Continuous statistics mode
|
Locating Method
|
Camera Locating
|
Color Space
|
CIE LAB,XYZ,Yxy,LCh,CIE LUV,HunterLAB
|
Color Difference Formula
|
ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00,ΔE(Hunter)
|
Other Colorimetric Index
|
WI(ASTM E313,CIE/ISO,AATCC,Hunter),YI(ASTM D1925,ASTM 313),TI(ASTM E313,CIE/ISO),
Metamerism Index MI, Staining Fastness,Color Fastness,Color Strength,Opacity
|
Observer Angle
|
2°/10°
|
Illuminant
|
D65,A,C,D50,D55,D75,F1,F2(CWF),F3,F4,F5,F6,F7(DLF),F8,F9,F10(TPL5),F11(TL84),F12(TL83/U30)
|
Displayed Data
|
Spectrogram/Values, Samples Chromaticity Values, Color Difference Values/Graph, Chromaticity Graph, Color Simulation, PASS/FAIL Result, Color Offset
|
Measuring Time
|
About 1.5s
|
Data Storage
|
Sample mode + QC mode: 40,000; Continuous statistics mode: 10,000
|
Repeatability
|
Spectral reflectance: Standard deviation within 0.08%
Chromaticity value: Within ΔE*ab 0.03 (When a white calibration plate is measured 30 times at 5 second intervals after white calibration)
|
Inter-instrument Error
|
Within ΔE*ab 0.2(Average for 12 BCRA Series II color tiles)
|
Measurement Mode
|
Single Measurement, Average Measurement(2 ~ 99 times)
|
Dimension
|
330*250*370mm
|
Weight
|
About 10 kg
|
Power Supply
|
DC 24V, 3A Power Adapter
|
Illuminant Life Span
|
5 years, more than 3 million times measurements
|
Display Screen
|
7" TFT Capacitive Screen-touch Display
|
Data Port
|
USB/RS-232, Bluetooth®4.0
|
Language
|
Simplified Chinese,Traditional Chinese,English
|
Working Environment
|
Temperature: 0~40℃; Humidity: 0~85% (No Condensation); Altitude < 2000m
|
Storage Environment
|
Temperature: -20~50℃; Humidity: 0~85% (No Condensation)
|
Standard Accessory
|
Power Adapter, User Manual, USB cable, Standard Calibration Board, Black Calibration Cavity,
|
Notes
|
The specifications are subject to change without notice.
|