| MEASURING RANGE |
0.01-4500 ppm
0.0004-200 mmol/l |
0.01-4500 ppm
0.0003-110 mmol/l |
0.01-4500 ppm
0.0014-600mmol/l |
0.50-4500 ppm
0.0125-110 mmol/l |
| DETECTION LIMIT |
0.01 ppm
0.0004 mmol/l |
0.01 ppm
0.0003 mmol/l |
0.01 ppm
0.0014 mmol/l |
0.03 ppm
0.0075 mmol/l |
| PRECISION |
0.4 % at 40 ppm
0.4 % at 1,74 mmol/l |
0.4 % at 40 ppm
0.4 % at 1,03 mmol/l |
0.4 % at 40 ppm
0.4 % at 5,71 mmol/l |
0.4 % at 40 ppm
0.4 % at 1,00 mmol/l |
| ACCURACY |
1 % at 40 ppm
1 % at 1.74 mmol/l |
1 % bei 40 ppm
1 % at 1.03 mmol/l |
1 % at 40 ppm
1 % at 5.71 mmol/l |
1 % at 40 ppm
1 % at 1.00 mmol/l |
| CALIBRATION |
Linear with 2 standards
Non linear with 6-8 standards,
cubic approximation |
| DRIFT |
1% in 60 min |
| REFERENCE |
Lithium guideline with 35 mg/l or 5 mmol/l
Caesium guideline on request |
| SAMPLE VOLUME |
2,5 ml |
| COMBUSTION GAS |
Propane, Butane or LPG: Recommended for alkaline elements
Acetylene for flame photometry: Recommended for alkaline earth elements |
| DISPLAY |
TFT-Display mit integriertem 8,4“ TFT Touchscreen, 800 x 600 Pixel |
| INTERFACES |
2 x USB
1 x Ethernet
1 x RS-232 for printer |
| EQUIPMENT |
Basic model +
Sampler and external sample table with control unit |
| ACCESSORIES |
24 character plain paper printer
Pressure reducing station with filter unit
Various standards and system solutions
Spare parts and accessories directly from the manufacturer |
| UPGRADEABILITY |