| MEASURING RANGE |
0.1-45000 ppm
0.004-2000 mmol/l |
0.1-45000 ppm
0.003-1100 mmol/l |
0.1-45000 ppm
0.014-6000 mmol/l |
5.0-45000 ppm
0.125-110 mmol/l |
| DETECTION LIMIT |
0.1 ppm
0.0004 mmol/l |
0.1 ppm
0.0003 mmol/l |
0.1 ppm
0.0014 mmol/l |
0.3 ppm
0.0075 mmol/l |
| PRECISION |
0.6 % at 40 ppm
0.6 % at 1.74 mmol/l |
0.6 % at 40 ppm
0.6 % at 1.03 mmol/l |
0.6 % at 40 ppm
0.6 % at 5.71 mmol/l |
0.6 % at 40 ppm
0.6 % at 1.00 mmol/l |
| ACCURACY |
1 % at 40 ppm
1 % at 1.74 mmol/l |
1 % at 40 ppm
1 % at 1.03 mmol/l |
1 % at 40 ppm
1 % at 5.71 mmol/l |
1 % at 40 ppm
1 % at 1.00 mmol/l |
| CALIBRATION |
UPGRADEABILITY |
| DRIFT |
1% in 60 min |
| REFERENCE |
Lithium guideline with 35 mg/l or 5 mmol/l
Caesium guideline on request |
| SAMPLE VOLUME |
0.25 ml |
| COMBUSTION GAS |
Propane, Butane or LPG: Recommended for alkaline elements
Acetylene for flame photometry: Recommended for alkaline earth elements |
| DISPLAY |
TFT-Display with integrated 8,4“ TFT touchscreen, 800x600 Pixel |
| INTERFACES |
2 x USB
1 x Ethernet
1 x RS-232RS-232 for printer |
| EQUIPMENT |
Basic unit +
External automation system, consisting of: Sampler, Diluter and Mixer |
| ACCESSORIES |
External automation system, consisting of: Sampler, Diluter and Mixer |
| UPGRADEABILITY |
Retrofit kits to other models available |