| Power Source |
120/240 V Selectable |
| Operation value |
Voltage: 63V
Current: 15A
Power: 200W |
| Constant Voltage |
Range: 1~63V
Accuracy: ±(1%+20mV)
Resolution: 20mV |
| Constant Current |
Range: 10mA~15A
Accuracy :
10mA~1A: ±(0.5%+1mA)
1A~6A: ±(0.5%+10mA)
6A~15A: ±(1%+100mA)
Resolution:
10mA~1A: 1mA
1A~15A: 10mA |
| Constant Power |
Range: 1~200W
Accuracy: ±1%+0.2W
Resolution: 0.1W |
| Constant Resistance |
Range: 0.5~999Ω
Accuracy: ±1%+0.2Ω
Resolution: 0.1Ω |
| Voltage Read back |
Accuracy: ±1%+50mV
Resolution: 20mV |
| Current Read back |
Accuracy: ±1%+10mA
Resolution: 10mA |
| Memory |
Save address: 1~9 |
| Timer |
Range: 1~999999 sec
Resolution: 1 sec |
| Fuse |
115V (100~120V): 1.0 A
230V (220~240V): 0.6 A |
| Protection |
Over Voltage & Over Current |
| Power Source |
120/240 V Selectable |
| Interface |
USB/RS232 |