| Measuring range | Cell constant | /cm | 0.01 | 0.1 | 1.0 | 
        
            | Electrical conductivity
 | μS/cm | 0.000 to 2.000 0.00 to 10.00
 | 0.000 to 2.000 0.00 to 20.00
 0.0 to 100.0
 | 0.0 to 200.0 0 to 1,000
 | 
        
            | mS/m | 0.0000 to 0.2000 0.000 to 1.000
 | 0.0000 to 0.2000 0.000 to 2.000
 0.00 to 10.00
 | 0.00 to 20.00 0.0 to 100.0
 | 
        
            | TDS conversion | mg/L | 0.00 to 2.00 0.0 to 10.0
 | 0.0 to 20.0 0 to 100
 | 0 to 200 0 to 1,0000
 | 
        
            | Electrical conductivity measurement and TDS conversion measurement cannot be selected at same time. | 
        
            | Temperature | ˚C | 0 to 100   Resolution: 0.01˚C | 
        
            | Repeatability | Electrical conductivity: ±0.5% full-scale or less TDS conversion: ±1.5% full-scale or less
 Temperature: ±0.1˚C (for equivalent input)
 | 
        
            | Linearity | Electrical conductivity: ±0.5% full-scale or less TDS conversion: ±1.5% full-scale or less
 Temperature: ±0.5˚C (for equivalent input)
 | 
        
            | Transmission output | 4 to 20 mA DC   Input/output insulated type Maximum load resistance 600 Ω
 | 
        
            | Contact input | One point   Contact format: Non-voltage a contact Contact function: Transmission output is held for closed contact input.
 | 
        
            | Temperature compensation range | 0 to 100˚C (but for 0˚C or less, 100˚C or more, extend and perform calculation) |