OPERATING ENVIRONMENT
• Temperature: 5ºC to 40ºC
• Relative humidity: 15% to 90%
DIMENSIONS
• W 41cm D 31cm H 21 cm
• Weight: 10.1 kg
INSTRUMENT CONTROL AND DISPLAY SOFTWARE
• Control Software: PDLiteProTM
• Analysis Software: PDViewTM
• Runs on any computer with Microsoft Windows XP or higher 32 bit operating system
• Computer-instrument communications via USB or Ethernet ports
PULSE ANALYSIS
• Shortest pulse width (FWHM): less than 2.5 ns, at the 3 dB cut-off magnitude
• Magnitude ranges:
– 20-340 mV in 20 mV steps,
– 50-850 mV in 50 mV steps,
– 100-1700 mV in 100 mV steps,
– 200-3400 mV in 200 mV steps
• Lowest magnitude window: 20 mV
• Magnitude windows: 32 per range
• Dual polarity
• 100 phase windows with 3.6º resolution
NOISE CANCELLATION
• Digital comparison of pulse arrival time and/or pulse shape, with magnitude difference up to 50%
• Resolution: 6 ns (corresponds to 2 m between a pair of sensors)
SIGNAL INPUT
• 12 BNC connectors for 80 pF epoxy-mica capacitors (multiplexor incorporated)
• 50 Ω input impedance
POWER FREQUENCY REFERENCE
• 50 Hz or 60 Hz
• Reference taken from input labelled “C1”
• Input range: 100 mV to 20 V
DATA ACQUISITION TIME
• Selectable at either 1 s or 5 s per magnitude window, for each coupler pair
POWER SOURCE
• 120/240 V, 50/60 Hz