GE Druck PACE 5000 General Specifications |
Pressure Measurement |
Standard Pressure Ranges |
25, 70, 200, 350 and 700 mbar gauge, 1, 2, 3.5, 7, 10, 20, 35, 70, 100, 135 and 210 bar gauge
0.35, 1, 3, 5, 10 15, 30, 50, 100, 150, 300, 500, 1000, 1500, 2000, 3000 psi
2.5, 7, 20, 35, 70, 100, 200, 350, 700 kPa, 1, 2, 3.5, 7, 10, 13.5, 21 MPa
All gauge versions available with negative calibration as standard. For absolute pressure ranges select a gauge range and add Barometric Reference (provides gauge + atmosphere absolute range). |
Over Range Indication |
10% above full scale pressure range. |
Pressure Media |
Dry, oil free, non-corrosive gas maintained at a value of 10% above the maximum required outlet pressure. Dry air or Nitrogen recommended. |
Display |
Panel |
¼ VGA wide format Graphics LCD, colour touch screen. |
Update Rate |
2 times per second. |
Readout |
± 9999999 |
Pressure Units |
24 scale units plus four user defined: Pa, hPa, kPa, MPa, kg/cm 2 , kg/m 2 , mmHg, cmHg, mHg, inHg, mmH 2 O, cmH 2 O, inH 2 O@4C, H 2 O@20C, H 2 O@60F, feetH 2 0 (@ 4C, 20C, 60F), psi, lb/ft 2 , torr, atm, mbar & bar. |
Performance |
Precision Ranges ≤ 200 mbar |
200 mbar= 0.02% Rdg + 0.02% FS, 70 mbar= 0.05% Rdg + 0.05% FS, 25 mbar= 0.1% Rdg + 0.1 % FS, includes linearity, hysteresis, repeatability and temperature effects for gauge pressure and assumes steady state and regular zeroing. For absolute pressures, add 0.1 mbar or 0.0015 psi |
Precision Ranges > 200 mbar |
0.01% Rdg + 0.01% FS includes linearity, hysteresis, repeatability and temperature effects for gauge pressures and assumes steady state and regular zeroing. For absolute pressures, add 0.1 mbar or 0.0015 psi. |
Negative Gauge Precision |
Maximum error at any given pressure value is equal to maximum error at the equivalent positive pressure value. |
Long Term Stability |
To 0.01% Reading per annum. 2 bar to 210 bar. 0.02% Rdg, 1 bar & 0.03% Rdg, 25 mbar-700 mbar |
Barometric Reference Precision |
Precision for the optional barometric reference 0.1 mbar or 0.0015 psi. Includes non-linearity, hysteresis, repeatability and temperature effects between 15°C (59°F) and 45°C (113°F). Long term measurement stability 0.1 mbar or 0.0015 psi per annum. |
Controller Stability |
0.003% FS |
Gas Consumption |
All supply gas is delivered to the system. No gas is used in measure mode, or when the instrument is turned off. |
Electrical |