Specifications |
|
9170 |
9171 |
9172 |
9173 |
Range (at 23 °C ambient) |
–45 °C to 140 °C
(–49 °F to 284 °F) |
–30 °C to 155 °C
(–22 °F to 311 °F) |
35 °C to 425 °C
(95 °F to 797 °F) |
50 °C to 700 °C†
(122 °F to 1292 °F) |
Display Accuracy |
± 0.1 °C full range |
± 0.1 °C: 35 °C to 100 °C
± 0.15 °C: 100 °C to 225 °C
± 0.2 °C: 225 °C to 425 °C |
± 0.2 °C: 50 °C to 425 °C
± 0.25 °C: 425 °C to 660 °C |
Stability |
± 0.005 °C full range |
± 0.005 °C: 35 °C to 100 °C
± 0.008 °C: 100 °C to 225 °C
± 0.01 °C: 225 °C to 425 °C |
± 0.005 °C: 50 °C to 100 °C
± 0.01 °C: 100 °C to 425 °C
± 0.03 °C: 425 °C to 700 °C |
Axial Uniformity (60 mm) |
± 0.1 °C at –45 °C
± 0.04 °C at –35 °C
± 0.02 °C at 0 °C
± 0.07 °C at 140 °C |
± 0.025 °C at –30 °C
± 0.02 °C at 0 °C
± 0.07 °C at 155 °C |
± 0.05 °C: 35 °C to 100 °C
± 0.1 °C: 100 °C to 225 °C
± 0.2 °C: 225 °C to 425 °C |
± 0.1 °C: 50 °C to 100 °C
± 0.25 °C: 100 °C to 425 °C
± 0.4 °C: 425 °C to 700 °C |
Radial Uniformity |
± 0.01 °C full range |
± 0.01 °C: 35 °C to 100 °C
± 0.02 °C: 100 °C to 225 °C
± 0.025 °C: 225 °C to 425 °C |
± 0.01 °C: 50 °C to 100 °C
± 0.025 °C: 100 °C to 425 °C
± 0.04 °C: 425 °C to 700 °C |
Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes) |
± 0.02 °C at –45 °C
± 0.005 °C at –35 °C
± 0.01 °C at 140 °C |
± 0.005 °C at –30 °C
± 0.005 °C at 0 °C
± 0.01 °C at 155 °C |
± 0.01 °C full range |
± 0.02 °C at 425 °C
± 0.04 °C at 700 °C |
Hysteresis |
0.025 °C |
0.04 °C |
0.07 °C |
Well Depth |
160 mm (6.3 in) |
203 mm (8 in) |
Resolution |
0.001 °C |
Display |
LCD, °C or °F, user-selectable |
Key Pad |