| Specifications | 
        
            | Test specifications | The accuracy specification for the display range is defined as ±(%reading + digit counts) at 23 °C ±5 °C, ≤75% RH. | 
        
            | Between 0 °C and 18 °C and between 28 °C and 40 °C, accuracy specifications may degrade by 0.1 x (accuracy specification) per °C. | 
        
            | The measurement range meets the service operating errors specified in EN61557-1: 1997, EN61557-2: 1997, EN61557-4: 1997, EN61557-6: 1997, DIN VDE0404-2. | 
        
            | Power-on test | The test indicates reversed L-N, missing PE, and measures the mains voltage and frequency. | 
        
            | Display range | 90 V to 264 V | 
        
            | Accuracy at 50 Hz | ±(2% + 3 counts) | 
        
            | Resolution | 1 V | 
        
            | Input impedance | > 1 MΩ // 2.2 nF | 
        
            | Maximum input mains voltage | 264 V | 
        
            | Earth bond test (R pe) | Display Range | 0 to 19.99 Ω | 
        
            | Accuracy (after bond test zeroing) | ±(2.5% + 4 counts) | 
        
            | Resolution | 0.01 Ω | 
        
            | Test current | 200 mA AC -0% +40% into 1.99 Ω 25 A AC ±20% into 25 mΩ at 230 V
 | 
        
            | Open circuit voltage | > 4 V AC, < 24 V AC | 
        
            | Bond test zeroing | can subtract up to 1.99 Ω | 
        
            | Insulation test (R iso) | Display range | 0 to 299 MΩ | 
        
            | Accuracy | ±(5% + 2 counts) from 0.1 to 300 MΩ | 
        
            | Resolution | 0.01 MΩ (0 to 19.99 MΩ) 0.1 MΩ (20 to 199.9 MΩ)
 1 MΩ (200 to 299 MΩ)
 | 
        
            | Test voltage | 500 V DC -0% + 10% at 500 kΩ load | 
        
            | Test current | > 1 mA at 500 kΩ load, < 15 mA at 0 Ω | 
        
            | Auto discharge time | < 0.5 s for 1 μF | 
        
            | Max. capacitive load | Operational up to 1 μF | 
        
            | Touch current test | Display range | 0 to 1.99 mA AC | 
        
            | Accuracy | ±(4% + 2 counts) | 
        
            | Resolution | 0.01 mA | 
        
            | Internal resistance (via probe) | 2 kΩ | 
        
            | Measuring method | Probe* | 
        
            | Substitute leakage current test | Display range | 0 to 19.99 mA AC | 
        
            | Accuracy | ±(2.5% + 3 counts) | 
        
            | Resolution | 0.01 mA | 
        
            | Test Voltage | 100 V AC ±20% | 
        
            | Load/leakage test: load current | Display range | 0 A to 13 A | 
        
            | Accuracy | ±(4% + 2 counts) | 
        
            | Resolution | 0.1 A* | 
        
            | Load/leakage test: load power | Display range 230 V Mains | 0.0 VA to 3.2 kVA | 
        
            | Accuracy | ±(5% + 3 counts) | 
        
            | Resolution | 1 VA (0 to 999 VA), 0.1 kVA (>1.0 kVA)* | 
        
            | Load/leakage test: leakage current | Display range | 0 to 19.99 mA | 
        
            | Accuracy | ±(4% + 4 counts) | 
        
            | Resolution | 0.01 mA* | 
        
            | PELV test | Accuracy at 50 Hz | ±(2% + 3 counts) | 
        
            | Overload protection | 300 V rms | 
        
            | Warning threshold | 25 V rms |