Specifications |
Test specifications |
The accuracy specification for the display range is defined as ±(%reading + digit counts) at 23 °C ±5 °C, ≤75% RH. |
Between 0 °C and 18 °C and between 28 °C and 40 °C, accuracy specifications may degrade by 0.1 x (accuracy specification) per °C. |
The measurement range meets the service operating errors specified in EN61557-1: 1997, EN61557-2: 1997, EN61557-4: 1997, EN61557-6: 1997, DIN VDE0404-2. |
Power-on test |
The test indicates reversed L-N, missing PE, and measures the mains voltage and frequency. |
Display range |
90 V to 264 V |
Accuracy at 50 Hz |
±(2% + 3 counts) |
Resolution |
0.1 V |
Input impedance |
> 1 MΩ // 2.2 nF |
Maximum input mains voltage |
264 V |
Earth bond test (Rpe) |
Display Range |
0 to 19.99 Ω |
Accuracy (after bond test zeroing) |
±(5% + 4 counts) |
Resolution |
0.01 Ω |
Test current |
200 mA AC -0% +40% into 1.99 Ω
10 A AC ±20% into 25 mΩ at 230 V |
Open circuit voltage |
> 4 V AC, < 24 V AC |
Bond test zeroing |
can subtract up to 1.99 Ω |
Insulation test (Riso) |
Display range |
0 to 299 MΩ |
Accuracy |
±(5% + 2 counts) from 0.1 to 300 MΩ |
Resolution |
0.01 MΩ (0 to 19.99 MΩ)
0.1 MΩ (20 to 199.9 MΩ)
1 MΩ (200 to 299 MΩ) |
Test voltage |
500 V DC -0% + 25% at 500 kΩ load or 250 V dc -0% +25% at 250 kΩ load |
Test current |
> 1 mA at 500 kΩ load, < 15 mA at 0 Ω |
Auto discharge time |
< 0.5 s for 1 μF |
Max. capacitive load |
Operational up to 1 μF |
Touch current test |
Display range |
0 to 1.99 mA AC |
Accuracy |
±(4% + 2 counts) |
Resolution |
0.01 mA |
Internal resistance (via probe) |
2 kΩ |
Measuring method |
Probe* |