Measurement Functions |
AC/DC Withstand Voltage, Insulation Resistance, Grounding,leakage, AC/DC Power |
Interface |
USB-HOST, RS232/RS485, HANDLER, ALARM, REMOTE |
Protect Function |
OV/OC |
Other Functions |
9‘’ LCD |
AC Withstand Voltage |
● |
Rate Ouput |
5kV AC,40mA |
Output Power |
200VA |
Output Voltage |
100 ~ 5000 V AC Resolution:resolution 1V |
Voltage Accuracy |
2% × Reading + 5Digits |
Output Frequency |
50Hz / 60Hz,Accuracy: ±0.1Hz |
Measurement Current |
0.005-9.999mA-20.00mA
Resolution:0.001mA/0.01mA |
Current Accuracy |
2% × Reading + 5Digits |
Rising Time |
0.1s-999.9s |
Testing Time |
0.5s-999.9s |
Falling Time |
0.1s-999.9s |
DC Withstand Voltage |
● |
Rate Output |
6kVdc/10mA |
Output Power |
120VA |
Output Voltage |
100~6kV DC
Resolution: 1V |
Voltage Accuracy |
2% × Reading + 5Digits |
Output Regulation |
<2% |
Rising Time |
0.1s-999.9s |
Testing Time |
0.5s-999.9s |
Falling Time |
0.1s-999.9s |
Insulation Resistance |
● |
Rate Output |
1kV/3GΩ |
Output Voltage |
100~1000V
Resolution:1V |
Voltage Accuracy |
2% × Reading + 5Digits |
Measurement Range |
0.1MΩ-3 GΩ |
IR Accuracy |
<500V:
0.1MΩ-100 MΩ: 5%×Reading+5Digits
100MΩ-1 GΩ : 10%×Reading+5Digits
≥500V:
0.1MΩ-100 MΩ: 2%×Reading+5Digits
100MΩ-2 GΩ: 5%×Reading+5Digits
2 GΩ-3 GΩ: 10%×Reading+5Digits |
Testing Time |
0.5s-999.9s |
Ground Impedance |
● |
Rate Output |
30A, 600mΩ |
Output Current |
30A |
Output Voltage |
<12V |
Frequency |
50Hz / 60Hz,Accuracy: ±0.1Hz |
Testing Time |
0.5s-999.9s |
Leak Current |
● |
Test Voltage Output |
Single-phase, requires external isolation transformer or variable frequency power supply to provide voltage and frequency required for testing |
Test Type |
G-L,G-N,AUTO(G-L,G-N) |
AC Voltage Measurement |
Range:5.0-300.0V
Resolution:0.1V
Accuracy: ±(1.5%×Reading+3V) |
AC Current Measurement |
Range:0.000-9.999-20.00mA
Resolution:0.001mA/0.01mA
Accuracy: 0.005-9.999mA ±(1.5%×Reading+0.005mA)
10.00-20.00mA ±(1.5%×Reading+0.05mA) |
Testing Time |
0.1s-999.9s |
AC Power |
● |
AC Voltage Output |
Single-phase, requires external isolation transformer or variable frequency power supply to provide voltage and frequency required for testing |
AC Voltage Measurement |
Range:0.0-300.0V
Resolution: 0.1V
Accuracy: ±(0.25%×Reading+0.25%×Range) |
AC Current Measurement |
Range:0.000-9.999-25.00A
Resolution:0.001A/0.01A
Accuracy: ±(0.25%×Reading+0.25×Range) |
Power Measurement |
Range:5.0-999.9-6000W
Resolution:0.1W/1W
Accuracy: ±(0.25%×Reading+0.25×Range) |
Power Factor |
Range:-1.00 – 1.000
Resolution:0.001
Accuracy: ±(0.5%×Reading+0.01) |
Testing Time |
0.1s-999.9s |
DC Power |
● |
DC Voltage Output |
Need external DC power supply to provide voltage and frequency required for testing |
DC Voltage Measurement |
Range:0.0-300.0V
Resolution: 0.1V
Accuracy: ±(0.25%×Reading+0.25%×Range) |
DC Current Measurement |
Range:0.000-9.999-25.00A
Resolution:0.001A/0.01A
Accuracy: ±(0.25%×Reading+0.25×Range) |
Power Measurement |
Range:5.0-999.9-6000W
Resolution:0.1W/1W
Accuracy: ±(0.25%×Reading+0.25×Range) |
Testing Time |
0.1s-999.9s |
Power Supply |
198V~242V AC
50Hz/60Hz
300VA |
Dimension & Weight |
400 x 200 x503.5(mm)
30kg |