Thiết bị phân tích Halogens & Sulfur AE-MIC YHS-11
Sample, that is set into the Auto-Sampler, is introduced into the combustion tube that carrier gas is clean air, and is combusted completely. Produced combustion gas is caught by the absorption unit, and is collected as the measuring liquid. Measuring liquid is automatically injected into the ion chromatograph device.
Respective halogen and sulfur in the sample is separately determined with the ion chromatograph, and is calculated as containing ratio of the sample.
- Ultra-micronization for organic trace element analysis
Ultra-micro-analyses (0.1 to 1.0mg) are attainable, as well as micro-analyses (1 to 10mg).
- Speeding up of analyses
5 to 6 samples can be measured per hour, including combustion and chromatogram development.
- Simultaneous multi-elements analysis
By oxygen/reduction-phase absorption liquid (patent), 5 elements (F, Cl, Br, I, S) can be simultaneously analyzed.
- Combustion standard containing multi-elements
In order to comply with organic calibration curve method regulated in JIS, high purified organic standard samples (NAC-st1, NAC-st2, NAC-st4, etc., patented) that contain multi-elements, are provided.
- Conformance to official analytical method
This analysis method is conformed to any kind of current official analytical method, such as JIS K0127
(Ion chromatograph general rule).
- Utilizing disposable sample boats
By utilizing inexpensive commercial magnetic boat (100JPY or so), residues after combustion are completely collected by the collection
function, and residues can be used for other analytical measurements,such as X-ray fluorescence, ICP-MS, etc.).
- Significant reduction of running cost
By utilizing clean air in place of cylinder gases (Ar, O2), the system can be moved discretionally,
without using expensive inert gases and oxygen gas.