Technical Data |
Oxygen concentration (O2) in flue gas |
Display |
Volume % referenced to dry flue gas |
Measurement principle |
Electrochemical sensor |
Range |
0.0-21.0 vol. % |
Accuracy |
± 0.3 Vol.-% |
Carbon monoxide (COV 10,000 ppm) in flue gas |
Display |
Volume ppm referenced to dry flue gas |
Measurement principle |
Electrochemical sensor, not H2 compensated |
Range |
0-10,000 vol. ppm; resolution 1 vol. ppm |
Accuracy |
±20 ppm (< 400 ppm), otherwise 10% of reading (if H2 < 5% of reading) |
Nitric oxide concentration (NOV) in flue gas (optional) |
Display |
Volume ppm referenced to dry flue gas |
Measurement principle |
Electrochemical Sensor |
Range |
0-3,000 vol. ppm, resolution 1 vol. ppm |
Accuracy |
±5 Vol.-ppm (<100 ppm), otherwise 5 % of reading |
Chimney draught / differential pressure (PD) (article 4520/4502) |
Display |
Pascal |
Measurement principle |
Semi-conductor diaphragm |
Range |
0.00 to ± 110.00 hPa; resolution 1 Pa |
Accuracy |
2 Pa (< 40 Pa), otherwise 5% of reading |
Flue gas temperature (TS) |
Display |
°F |
Measurement principle |
Thermocouple (NiCr-Ni) |
Range |
68-1472 °F |
Accuracy |
0-271.4 °F
271.4-1472 °F: ± 1.5 % of reading |
Combustion air temperature (TA) |
Display |
Fahrenheit |
Measurement principle |
Thermocouple (NiCr-Ni) |
Range |
68-212 °F |
Fahrenheit |
±1°F |
Power supply |
Wohler A 450 L |
Lithium-Ion, rechargeable battery 3.7 V, 2,250 mAh, charges via USB |
Battery operating time |
Wohler A 450 L |
approx. 6 h (depends on operating status and display illumination) |
Charging time |
Wohler A 450 L |
approx. 3 h |
Storage temperature |
-4-122 °F |
Operating temperature |
41-104 °F |
Weight |
1.5 lbs. |
Dimension |
6x4x1.5 inches |
Length of the probe hose |
5 ft. |