The dLW139.153-75D6 Flexure Table system combines our ET-139 permanent magnet field shaker and dPA-153 amplifier together with our 6 inch square Flexure suspension Table. This combination provides a large mounting surface with increased load carrying capacity for larger or heavier than normal test specimens not requiring the force of a larger shaker system. The unique flexure suspension table guidance eliminates the requirement for the oil found in slip table systems. This oil-less design allows operation at other than the horizontal orientation, and adds increased suspension guidance reducing the effects of heavy and off-center loads on the shakers armature suspension system.. The Table flexures are fully rated for 1 inch p-p operation, and the shaker can be uncoupled from the table for operation in its normal vertical orientation with specimens mounted directly to the shakers armature.
The air cooled dPA-153 dLAB shaker amplifier delivers full system performance while minimizing power and amplifier cooling requirements. The amplifier's multi-function digital display provides amplifier operational status as well as system peripheral and complete vibration system operational diagnositcs. Built in shaker-specific current limiting and full internal, system and external electrical interlock connections and logic help protect the system and test specimen from accidental abuse.
|