AC Hipot test
|
|
Booster rated output capacity
|
AN9640A(F)
|
500VA(5000V/100mA)1
|
|
AN9640B(F)
|
100VA(5000V/20mA)
|
|
Booster maximum short-circuit current
|
>200mA
|
|
Output waveform
|
Sine wave, <3%(pure resistance,5000V/1mA),<3%(From empty load to full load)
|
|
distortion and regulation
|
|
Output voltage setting
|
Range
|
500~5000V
|
|
Resolution
|
5V
|
|
Accuracy
|
±(3%×setting+10V)
|
|
Frequency setting
|
Range
|
50Hz/60Hz
|
|
Accuracy
|
±0.1%×setting
|
|
Alarming limits setting
|
Current upper-limit
|
0.10~99.99mA(20.00mA for AN9640B/C(F))
|
|
Current lower-limit
|
0.00~10.00mA
|
|
Compensation current setting
|
Range
|
0.00~10.00mA, automatic testing, compensation can be set to ON/OFF
|
|
Time setting
|
Range
|
1~999s
|
|
Ramp up time range
|
1~100s
|
|
Accuracy
|
±(0.1%×setting+1count)
|
|
Breakdown current alarm
|
Range
|
0.10~99.99mA(20.00mA for AN9640B/C(F))
|
|
Accuracy
|
±(3%×reading+3counts)
|
|
Insulation Resistance test
|
|
Rated output capacity
|
2VA(1000V/2mA)
|
|
Output voltage regulation
|
<3%(From no load to full load)
|
|
Output voltage ripple
|
<1%
|
|
Output voltage setting
|
Range
|
DC 500Vor1000V
|
|
Accuracy
|
±(3%×setting+3counts)
|
|
Alarming limits setting
|
Resistance upper-limit range
|
0.0~99.9 MΩ,100~2000MΩ(If upper-limit is set to 0, it will not make adjudgment for upper limit )
|
|
Resistance lower-limit range
|
0.3~99.9 MΩ, 100~2000MΩ
|
|
Time setting
|
Range
|
1~999s
|
|
Accuracy
|
±(0.1%×setting+1count)
|
|
Insulation resistance measurement
|
Range
|
0.0~99.9 MΩ, 100~2000 MΩ
|
|
Accuracy
|
≤200MΩ: ±(3%×reading+3counts)
|
|
|
Remarks:Except 500V and 1000V voltage points, other voltage accuracy are all 10%. Insulation resistance accuracy is under temperature 0~40℃ , humidity 40℃/(20~75)% RH,otherwise the accuracy will change.
|
|
Ground Bond Resistance test
|
|
No load output voltage
|
<12V
|
|
Output current setting
|
Range
|
3~30A
|
|
Accuracy
|
±(3%×setting+0.2A)
|
|
Current frequency setting
|
Range
|
Sine waveform 50Hz/60Hz
|
|
Accuracy
|
±0.1%×setting
|
|
Test time setting
|
Range
|
1~999s
|
|
Accuracy
|
±(0.1%×setting+1count)
|
|
Result display mode
|
Ground resistance or voltage
|
|
Ground resistance display mode:
|
|
Alarming limits setting
|
Resistance upper-limit range
|
3A≤output current≤10A:10~600 mΩ
|
|
|
|
Resistance lower-limit range
|
0~100 mΩ
|
|
Compensation resistance setting
|
Range
|
0~200 mΩ, automatic testing, compensation can be set to ON/OFF
|
|
Ground bond resistance measurement
|
Range
|
10~600 mΩ
|
|
Accuracy
|
±(3%×reading+ 5counts)
|
|
Ground voltage display mode:
|
|
Alarming limits setting
|
Voltage upper-limit range
|
30~7500 mV
|
|
|
Voltage lower-limit range
|
0~3000 mV
|
|
Compensation voltage setting
|
Range
|
0~1000 mV, automatic testing, compensation can be set to ON/OFF
|
|
Ground bond resistance measurement
|
Range
|
10~7500 mV
|
|
|
Accuracy
|
±(3%×reading+ 50mV)
|