AC WITHSTANDING (all models) |
Output-Voltage Range |
0.100kV~5.000kV ac |
Output-Voltage Resolution |
2V/step |
Output-Voltage Accuracy |
± (1% of setting + 5V) [no load] |
Maximum Rated Load |
500 VA (5kV/100mA) |
Maximum Rated Current |
100mA (0.5kV< V 5kV)
10mA (0.1kV V 0.5kV) |
Output-Voltage Waveform |
Sine wave |
Output-Voltage Frequency |
50 Hz / 60 Hz selectable |
Voltage Regulation |
± (1% of reading + 5V) [full load → no load] |
Voltmeter Accuracy |
± (1% of reading + 5V) |
Current Measurement Range |
0.001mA~100.0mA |
Current Best Resolution |
0.001mA / 0.01mA /0.1mA |
Current Measurement Accuracy |
±(1.5% of reading + 30 counts) when HI SET < 1.11mA
±(1.5% of reading + 3 counts) when HI SET 1.11mA |
Window Comparator Method |
Yes |
ARC Detect |
Yes |
Sweep Function* |
Yes |
RAMP (Ramp-Up Time) |
0.1s~999.9s |
TIMER (Test Time)* |
OFF, 0.5s~999.9s |
GND |
ON/OFF |