The operating power supply |
AA batteries (1.5V) 8 This
battery mark at the time of the exchange of * battery display
external power supply DC12V |
ZCT measurement unit |
Measurement range |
20/200 / 2000mA (detecting secondary AC voltage ZCT) |
Display |
4-digit 1/2 LCD digital display |
Measurement system |
Effective value conversion method of the mean value measurement |
Input impedance |
40kΩ |
Measurement accuracy |
± 5% rdg ± 10dgt (10 % ~ 100% of each range)
in the following grounding current 200mA, if you make a gain correction. |
Sample rate |
2.5 times / sec. |
Over Display |
Only one display of maximum digit numeric |
Gain alternating current |
Output current |
15mA 20Hz (sine wave) |
Load impedance |
30Ω |
Output accuracy |
± 0.5mA |
Operating temperature limit |
5 ~ 40 ℃ 20 ~ 80% RH |
External dimensions |
130 (D) × 183 (W) × 100 (H) |
weight |
1kg (including battery) |
accessories |
Measurement code, the terminal code (alligator type), terminal code (Round terminal type) |
option |
Terminal code (Round terminal type) standard price (consumption tax not included)
AC adapter standard price (consumption tax not included) |