Use power |
AC100V 50 / 60Hz ± 10% |
Voltage element output |
|
DC5 ~ 750V (continuously variable) |
Output current |
500mA |
Output switching |
Test switch "ON" second-hand voltage and current output
* 64PB output operation time measurement teeth, 0V scolded test voltage similar sudden change to another time wo measured masu. (P element output) |
Current element output |
|
DC5 ~ 120V (continuously variable) |
Output current |
10A (continuous rating)
15A (5 min rating) maximum output voltage of 100V |
Output switching |
Voltage and current output at the test switch "ON"
* 64PB output operation time measurement, we can measure the time that a sudden change in the test voltage from 0V. (B element output) |
Output resistance switching |
Resistance switching for current adjustment
and high range 20Ω low range 5Ω short 0Ω
high-range 5A output (5 min rating)
low range 10A output (5 min rating) 15A output (1 minute rating)
short circuit 10A (continuous rating) 15A (5 min rating ) |
Voltage element voltmeter |
|
750.0V (resolution 0.1V) 7 segment LCD |
Measurement accuracy |
± 0.5% rdg ± 5dgt |
Sampling rate |
2.5 times / sec. |
Current element voltmeter |
|
199.9V (resolution 0.1V) 7 segment LCD |
Measurement accuracy |
± 0.5% rdg ± 5dgt |
Sampling rate |
2.5 times / sec. |
Current element ammeter |
|
19.99A (resolution 0.01A) 7 segment LCD |
Measurement accuracy |
± 0.5% rdg ± 5dgt |
Sampling rate |
2.5 times / sec. |
counter |
Measurement range |
0 ~ 199.999sec resolution 01Ms
200.00 ~ 1999.99Sec resolution 10ms (automatic carry) |
Measurement accuracy |
0.01% rdg ± 1dgt ± 5ms ± ⊿ t ⊿ t:
stop signal Niyoru each error
contact, DC voltage
± voltage
(05 voltage
(10 voltage (20V ERROR) ± 1ms |
Stop signal |
Contact a contact, b contact automatic detection
voltage direct current, alternating both 10 ~ 220V applied, removed |
Timer (B, P element common) |
Set time |
0.001s ~ 9999h |
Number of digits |
4-digit |
Output time error |
Maximum 30ms |
Output contact (B, P element each output) |
Contact capacity |
AC / DC110V 0.1A |
External dimensions |
662 (W) × 247 (D) × 440 (H) |
weight |
About 23kg |
accessories |
Lead (power cord, auxiliary power cord, current element output code, voltage element output code, timed measurement code, earth code)
lead storage bag
photo MOS relay
manual |