Fulfills Standards
|
IEC 61672 Class 2
|
IEC 61260 Class 2
|
IEC61252:2002
|
Microphone
|
1/2” prepolarized condenser microphone for free field (Sensitivity Level: -36dB)
|
Preamplifier
|
AWA14602 removable preamplifier
|
Frequency Range
|
20 Hz ~ 12.5kHz ± 1 dB (not including microphone)
|
Total Measurement Range
|
30dB-133dB
|
Frequency Weighting.
|
Parallel (simultaneous) A, C, Z, B, D
|
Time Weighting
|
Parallel (simultaneous) F, S, I, Peak
|
Measurement Parameters
|
Lxyp, Lxyi, Lxeq,1s, Lxeq,T, LAE, E, Cpeak, LAFmax, LAfmin, LAfeqT, LAseqT, LAieqT.
Note: X is A, C, Z and Y is F, S, I
|
A/D Bits:
|
24 bits
|
Sampling Frequency
|
32 kHz.
|
Calibration
|
Using Sound Calibrator Class 2 model AWA6221B
|
Correction Function
|
Diffusion field correction in order to comply with standards ANSI S1.4
|
Display
|
240×320 color screen, adjustable brightness, backlight can be closed
|
Display Resolution
|
0.1 dB
|
Low battery indication
|
indicate low battery
|
Data Storage
|
3328 groups of integrating measuring results only.
32 Mb FLASH RAM
(32G SD card memory function is optional)
|
Print
|
Mini-printer
|
Internal Clock
|
Error less than 1 min/month
|
Output Interface
|
AC Output (full scale): 1.0V AC RMS; Output Impedance: 1k Ω; Connector: φ3.5 mm stereo plug
|
DC Output: 20mV/dB; Output Impedance: 1k Ω; Connector: DB-9 plug
|
RS232 Interface: To computer for output some measurement results instantaneous values , also to mini-printer for printing
Transmission speed: 4800, 9600,115200 bps
|
USB Interface: available and no need device drive.
|
Power Supply
|
4×LR6 alkaline battery or rechargeable batteries
|
5 V external power supply
|
Battery Life
|
Longest time of 30 hours continuously with 4×LR6 alkaline battery
|
Dimensions
|
240 (H) x 80 (W) x 30 (D), mm.
|
Weight
|
0.35 kg. (include batteries)
|
EMC
|
Type X
|
Environment:
|
Working Temperature: -10 ~ 50 °C
Storing Temperature: -20 ~ 70 °C
Relative Humidity: 25 ~ 90 %
|