Fulfils Standards
|
IEC 61672-2013 Class 1
|
IEC 61260-2014 Class 1
|
ANSI S1.11: 2004
|
IEC 60651:2001 Type 1
|
IEC 60804: 2000 Type 1
|
IEC61252:2002
|
Microphone+preamplifier
|
1/2” prepolarized condenser microphone+ Preamplifier
(Sensitivity Level: -28dB)
|
Preamplifier
|
AWA14601 removable preamplifier type 1
|
Frequency Range
|
10 Hz ~ 20 kHz ± 1 dB (not including microphone)
|
Total Measurement Range
|
Low range: (20~132)dBA, (25-132)dBC, (30-132)dBZ
High range: (30~142)dBA, (35-142)dBC, (40-142)dBZ
|
Self-generated Noise
|
<12 dB(A), 17 dB(C), 22 dB(Z)
|
Frequency Weighting.
|
Parallel (simultaneous) A, C, Z, B, D and user1&2-defined weighting
|
Time Weighting
|
Parallel (simultaneous) F, S, I, Peak
|
Measurement Parameters
|
Lxyp, Lxyi, Lxeq,1s, Lxeq,T, LAE, E, Cpeak, LAFmax, LAfmin, LAfeqT, LAseqT, LAieqT.
Note: X is A, C, Z and Y is F, S, I
|
A/D Bits:
|
24 bits
|
Sampling Frequency
|
48 kHz
|
Calibration
|
Using Sound Calibrator Class 1 : IEC60942:2003 / IEC 60942:2017
model AWA6221A
|
Correction Function
|
Press field correction in order to comply with standards ANSI S1.4
|
Delay Time
|
The meter can delay 0~99s after pressing start measuring button
|
Back Erase Function
|
Elimination of undesired noise; example barking dogs, cars, doors
|
Display
|
240×320 color screen, adjustable brightness, backlight can be closed
|
Display Resolution
|
0.1 dB
|
Low battery indication
|
Symbol indicate low battery
|
Data Storage
(32 Mb FLASH RAM.
SD memory card is optional)
|
- Ø 3328 groups of integrating measuring results only.
- Ø 3328 groups of statistical results only (‘statistical 1’ and ‘statistical 2’ analysis index are same.)
- Ø 2663 groups of statistical results only (‘statistical 1’ and ‘statistical 2’ analysis index are different.)
|
Print
|
Mini-printer
|
Internal Clock
|
Error less than 1 min/month
|
Output Interface
|
AC Output (full scale): 1.0V AC RMS; Output Impedance: 1k Ω; Connector: φ3.5 mm stereo plug
|
DC Output: 20mV/dB; Output Impedance: 1k Ω; Connector: DB-9 plug
|
RS232 Interface: To computer for output some measurement results instantaneous values , also to mini-printer for printing
Transmission speed: 4800, 9600,115200 bps
|
USB Interface: available and no need device drive.