Laminar Flow Tube Flow Range |
10, 20, 50, 100, 200, 500 mL/min
1, 2, 5, 10, 20, 30, 50, 100, 200, 500 L/min |
Pressure Media |
Air |
Accuracy |
±2% of F.S. ±1 digit |
Conversion Temperature / Atmospheric Pressure |
20℃ (or 0℃)
Flow rate is based on 1 atmospheric pressure |
Operating Temperature |
5 to 35℃ |
Test Pressure |
Depends on the selected flow ranges |
Temperature Sensor |
Built in the Laminar Flow Tube or connected to the outlet of Laminar Flow Tube |
Generated Differential Pressure |
Approx. 0.6 kPa at F.S. flow |
Pressure Loss |
150% of Measured Differential Pressure |
Port Size |
Differential Pressure Connection Port: Rc 1/8
Flow Sensor Connection Port: Refer to “Laminar Flow Tube Flow Ranges” on the next page |
DPS Proof Pressure |
1 MPa (PT-110FC-A), 10 kPa (PT-103B-A) |
Power source |
24 VDC ± 10%, 0.3A max With A/C Adaptor |
External Dimensions |
Display: W140 × H66 × D172 mm |
Panel Cutting Dimensions |
W133 x H61 mm |
Documents |
Traceability documents, Operation manual |