| Flow range (Display value)
 | 500mL/min (000~500) 2L/min (0.000~2.000) 5L/min (0.00~5.00) 20L/min (00.00~20.00) 50L/min (00.0~50.0) 100L/min (000.0~100.0) | 
        
            | Flow sensor | Mass flow sensor | 
        
            | Pressure media | Air | 
        
            | Accuracy | ±1.5% of F.S. ±1 digit (At specified pressure) ±3% of F.S. ±1 digit (At non-specified pressure) | 
        
            | Specified test pressure range | Micro (L03) : 10~30kPa Low (L1) : 30~80kPa Medium (M) : 30~700kPa Vacuum (V) : -10~-70kPa | 
        
            | Pressure proof | 1.5MPa | 
        
            | Channels | 32ch (0~31) | 
        
            | Timer setting | 999.9 sec (min 0.1sec) | 
        
            | Power supply | AC100~240V±10%, 50/60Hz,100VA max (Power cable included is for AC125V only)
 | 
        
            | Pressure source | Clean and dry air. Pressure source should be higher than the test pressure | 
        
            | Connecting port | Rc 1/4 (Pressure source, Pilot pressure) Rc 1/2 (WORK port) | 
        
            | CPU | 16 BIT CPU | 
        
            | Ambient temperature | 5~45 ℃ | 
        
            | Humidity | Max. 80%RH and no condensation | 
        
            | Weight | Approx. 12kg | 
        
            | RS-232C Format | T Format  Fixed length output | Output for judgment signal and flow rate data | 
        
            | Serial printer output | Printer format | (Data output) Judgment result, Leak rate, Compensation value, Test pressure, WK#, CH#, Time etc. (Setting output) Setting parameter of all screen | 
        
            | Flow rate display | L/min、mL/min、L/s、mL/s、L/h、m3/h、 USP (User Span) | 
        
            | Pressure Unit | kPa、MPa | 
        
            | Standard and accessories | Quick connection fittings, I/O connector, Power cable (3m), Inspection record, Traceability documents and Operation manual 
 
 |