Pressure media |
Air |
Connecting port |
Pressure source, Pilot pressure:Rc 1/4
Laminar Flow model 100L only: Rc 3/8
WORK port:200L/min or less: Rc 1/4
500L/min or less: Rc 1/2
|
Accuracy |
*Laminar Flow Sensor
±1.5% of F.S. ±1 digit (At specified pressure) |
*Mass Flow Sensor
±1.5% of F.S. ±1 digit (At specified pressure)
±3% of F.S. ±1 digit (At non-specified pressure) |
CPU |
ARM9 400 MHz, DRAM 128 MB |
RS-232 |
Front panel port |
Fixed length output: T, IL, ML, D, P |
Specified test pressure range |
Micro (L01): 1 to 10kPa (Regulator is not enclosed)
Micro low (L03): 10 to 30kPa (Mass Flow only)
Micro low (L05): 10 to 50kPa (Laminar Flow only)
Low (L): 30 to 80kPa
Medium (M): 3to 700kPa
Vacuum (V): -10 to -70kPa (20L or less) |
Rear panel port |
Fixed length output: T, IL, ML, D, P |
Test Data |
Flow, Pressure, Comp value, Air temp, Flow limits, Atm press, and others |
CSV Copy to USB |
CSV file |
USB Port |
Parameter backup
System backup
Software version update
Copy Operation manual (PDF) |
Channels |
32ch (0~31) |
Power supply |
AC100 to 240V ±10%, 50/60Hz, 60VA max
(Power cable included is for AC125V only) |
Flow Unit |
L/min, mL/min, L/s, mL/s, L/h, m3/h, USP(User Span) |
Timer setting |
999.9 sec (min 0.1sec) |
Pressure Unit |
kPa, MPa, (psi, kg/cm2, bar, mbar, mmHg, cmHg, inHg, mmH2O) |
Pressure source |
Clean and dry air. Pressure source should be higher than the test pressure |
Standard and accessories |
Quick connection fittings, I/O connector, Power cable (3m), Inspection record, Traceability documents and Operation manual |
Ambient temperature |
5 to 45℃ |
Humidity |
Max. 80%RH and no condensation |
|
|
Weight |
Approx. 15kg
|